The Apera Instruments PH60 Pocket pH Tester is designed for reliable and easy pH measurement in general water solutions. This includes applications such as hydroponics, horticulture, aquaculture, pools & spas, water treatment, cooling towers, environmental monitoring, beverage making, and educational settings. The meter features a replaceable pH probe (PH60-E) specifically for general water solutions. It is also compatible with three other types of pH electrodes and an ORP probe (sold separately) for specialized applications like solid samples and surface tests.

Athugið: A few drops of 3M KCL liquid (included in the kit) should be added to low ion-strength test samples (e.g., distilled water or RO water) to significantly accelerate stabilization without affecting the pH reading.

Close-up of the PH60 pH sensor bulb

Mynd: Nærmynd view of the high-quality glass pH sensor bulb, designed for accuracy and stability.

Infographic showing various applications for the PH60 pH tester

Image: Visual representation of diverse applications including hydroponics, pools, brewing, drinking water, water treatment, aquariums, and aquaculture.

2. Hvað er í kassanum

Your Apera Instruments PH60 Waterproof pH Pocket Tester Kit includes the following items:

Apera Instruments PH60 pH Tester Kit contents laid out in its carrying case

Image: The complete PH60 pH Tester Kit, showing the meter, calibration solutions, bottles, and accessories neatly organized in its hard-shell case.

3. Uppsetning

3.1 Uppsetning rafhlöðu

The PH60 tester requires 4 AAA batteries (included). To install or replace batteries:

  1. Unscrew the battery cap at the top of the meter.
  2. Settu tvær AAA rafhlöður í og ​​gætið þess að pólarnir (+/-) séu rétt pólaðir.
  3. Skrúfið rafhlöðulokið örugglega aftur á.
PH60 pH tester with battery compartment open, showing AAA batteries

Image: The top section of the PH60 pH tester with the cap removed, revealing the battery compartment for AAA batteries.

3.2 Upphafleg undirbúningur rannsakanda

Before first use or if the tester has not been used for a long time (over 1 month):

  1. Pull out the battery insulation slip (if present) and remove the probe cap.
  2. Rinse the probe thoroughly with pure water (preferably distilled or deionized water, RO water is acceptable). Shake off excess water.
  3. If the tester hasn't been used for a long time (over 1 month), soak the probe in the 3M KCL soaking solution (included in the kit) for 15 minutes.
  4. Proceed to calibrate the meter before testing.

Video: Apera Instruments PH60 pH Tester Quick Start Tutorial. This video demonstrates the initial preparation steps, including removing the battery insulation slip, rinsing the probe, and soaking it in KCL solution if necessary (0:13 - 0:55).

4. Notkunarleiðbeiningar

4.1 pH kvörðun

The PH60 tester features 1-3 point automatic calibration with auto-buffer recognition. Always calibrate pH 7.00 first. If you need to calibrate pH 4.00 or pH 10.01, do so immediately after pH 7.00 calibration. Do NOT turn the meter off between pH 7.00 and pH 4.00/10.01 calibrations.

  1. Pour pH buffer solutions (pH 4.00, pH 7.00, pH 10.01) into the corresponding calibration vials to about half volume.
  2. Skolaðu mælinn í hreinu vatni og hristu umfram vatn af.
  3. Short press the power button to turn on the tester.
  4. Dip the probe into the pH 7.00 buffer solution. Make a quick stir, then let it stand.
  5. Long press the 'CAL' button to enter calibration mode. The screen will turn green.
  6. Wait for the reading to stabilize (a smiley icon will appear on the screen), then short press 'CAL' to finish the first point calibration. The tester will return to measurement mode, and an 'M' icon will appear at the bottom left, indicating a successful first-point calibration.
  7. Without turning off the tester, rinse the probe and repeat steps 4-6 for pH 4.00 buffer solution. An 'L' icon will appear for successful 2-point calibration.
  8. Without turning off the tester, rinse the probe and repeat steps 4-6 for pH 10.01 buffer solution. An 'H' icon will appear for successful 3-point calibration.
Step-by-step guide on how to calibrate the PH60 pH tester

Image: Detailed infographic illustrating the 3-point calibration process for the PH60 pH tester.

Video: Apera Instruments PH60 pH Tester Quick Start Tutorial. This segment demonstrates the pH calibration process (0:56 - 4:26).

4.2 pH-mæling

To take a pH measurement:

  1. Short press the power button to turn on the tester.
  2. Skolaðu mælinn í hreinu vatni og hristu umfram vatn af.
  3. Insert the probe into your sample solution, make a quick stir, and hold still.
  4. Take the reading as the measurement when it is fully stabilized (a smiley icon appears and stays on screen).
  5. If the Auto-Hold function is turned on (in parameter setting P4), the reading will be automatically locked when it's stable for more than 10 seconds. The 'HOLD' icon will stay on. Short press the 'CAL' button to cancel Auto-Hold and continue measuring.
  6. Short press the 'MODE' button to switch to mV mode, which is for ORP testing or technical analysis (an ORP probe is needed for ORP measurements).
Hand holding the PH60 pH tester, showing a pH reading of 7.00

Image: A hand holding the PH60 pH tester, displaying a pH reading of 7.00 and a temperature of 25.0°C.

Video: Apera Instruments PH60 pH Tester Quick Start Tutorial. This segment demonstrates the pH measurement process (4:26 - 5:33).

5. Viðhald

5.1 Probe Cleaning

The accuracy of the tester depends on the cleanliness of the probe. Always thoroughly rinse off the probe before and after each measurement with pure water.

Video: Apera Instruments PH60 pH Tester Quick Start Tutorial. This segment demonstrates probe cleaning procedures (5:33 - 6:36).

5.2 Probe Storage

Proper storage is crucial for maintaining probe longevity and accuracy.

IMPORTANT: NEVER store the probe in pure water (like tap, RO, distilled, or deionized water) as they could damage the pH sensor. If this happens, immediately soak the pH probe in the 3M KCL soaking solution overnight to restore. Then re-calibrate it before using. Pure water is only for rinsing off the probe, NOT for storage.

Video: Apera Instruments PH60 pH Tester Quick Start Tutorial. This segment demonstrates proper probe storage (6:36 - 7:29).

6. Bilanagreining

Hér eru nokkur algeng vandamál og lausnir þeirra:

7. Tæknilýsing

EiginleikiGildi
Vörumál7 x 1.6 x 1.6 tommur
Þyngd hlutar10 aura
Tegundarnúmer vöruAI311
Rafhlöður4 AAA rafhlöður nauðsynlegar (fylgir með)
pH nákvæmni±0.01 pH
pH svið-2.00 til 16.00 pH
Hitastig32 til 122°F (0 til 50°C)
Sjálfvirk ræsing8 mínútur án aðgerða

8. Ábyrgð og stuðningur

Apera Instruments is committed to providing reliable products and ongoing support to ensure you achieve the most accurate results. For warranty information, technical assistance, or any product-related inquiries, please refer to the official Apera Instruments website or contact their customer support directly. Contact details are typically provided on the product packaging or within the full user manual.

Tengd skjöl - AI311

Preview Notendahandbók fyrir Apera PH60 pH-mæli - Leiðbeiningar og upplýsingar
Ítarleg notendahandbók fyrir Apera Instruments PH60 Premium pH-mælitækið, sem fjallar um uppsetningu, kvörðun, mælingar, bilanaleit og upplýsingar.
Preview Apera EC20 vasaleiðnimælir: Leiðbeiningar og leiðbeiningar
Ítarleg leiðbeiningarhandbók fyrir Apera EC20 vasaleiðniprófarann. Kynntu þér uppsetningu, kvörðun, leiðnimælingar, tæknilegar upplýsingar og ábyrgðarupplýsingar.
Preview Notendahandbók fyrir Apera Instruments PH20 pH-mælibúnað
Ítarleg notendahandbók fyrir Apera Instruments PH20 pH mælitækið, þar á meðal uppsetning rafhlöðu, virkni takkaborðsins, kvörðun, mælingar, stillingar breytu, bilanaleit, tæknilegar upplýsingar og ábyrgðarupplýsingar.
Preview Notkunarhandbók fyrir Apera Instruments 5052 vasamælingu á seltu í sjó
Ítarleg notkunarhandbók fyrir Apera Instruments 5052 vasamæli fyrir söltun sjávar, sem fjallar um tæknilega þætti, kvörðun, mæliaðferðir, mikilvægar athugasemdir og ábyrgðarupplýsingar.
Preview Notendahandbók fyrir Apera PH60-HT pH-mæli: Háhita- og ætandi lausnir
Ítarleg notendahandbók fyrir Apera PH60-HT pH-mælinn, hannaður fyrir nákvæmar pH-mælingar við háan hita (allt að 140°C) og ætandi lausnir. Fjallar um innihald búnaðarins, notkun, kvörðun, umhirðu rafskauta, stillingar breytna, tæknilegar upplýsingar, bilanaleit og ábyrgð.
Preview Notendahandbók fyrir Apera Instruments PH60 Premium pH-mæli
Ítarleg notendahandbók fyrir Apera Instruments PH60 Premium pH-mælitækið, sem fjallar um uppsetningu, kvörðun, mælingar, viðhald, bilanaleit og tæknilegar upplýsingar.